Page 1
INSTRUMENTAL METHODS OF ANALYSIS
3.1 INTRODUCTION
Analytical instrumentation plays an important role in the production and evaluation of new
products and in the protection of consumers and environment. It is used in checking the quality
of raw materials such as substances used in integrated circuit chips, detection and estimation of
impurities to assure safe foods, drugs, water and air, process optimization and control, quality
check of finished products and research and development. Most of the modern instruments are
microprocessor/computer controlled with user friendly software for collection of data, analysis
and presentation.
This chapter deals with the different types of analytical instrumental methods that find use in a
variety of industries. These include molecular spectroscopic methods, thermal methods of
analysis, X-ray diffraction, scanning electron microscope and sensors.
3.2 SPECTROSCOPY
It is the study of interaction of electromagnetic radiation with matter consisting of atoms and
molecules. When a substance is irradiated with electromagnetic radiation, the energy of the
incident photons may be transferred to atoms and molecules raising their energy from ground
state level to excited state. This process is known as absorption and the resultant spectrum is
known as absorption spectrum. The process of absorption can occur only when the energy
difference between the two levels E is exactly matched by the energy of the incident photons as
given by the equation
E = h? = hc/?
where h is Planck’s constant(6.63 x 10
-34
Js), ? is the frequency of incident radiation, c is the
velocity of light and ? is the wavelength of the incident radiation. The excited state atoms and
molecules then relax to the ground state by spontaneous emission of radiation. The frequency of
the radiation emitted depends on E.
The energy changes that occur in atoms and molecules during interaction with different regions
of electromagnetic radiation are given below.
Radiation
absorbed
Energy of the
radiation
(J/mole)
Effect on the
atoms/molecules
Applications
Introduction, absorption of radiation, UV-Visible Spectrophotometer: Instrumentation and
application, IR Spectrophotometer: Instrumentation and application, Thermal methods of analysis-
TGA, DTA, DSC, Sensors: Oxygen and Glucose sensor, Cyclic Voltammetry for redox system.
Page 2
INSTRUMENTAL METHODS OF ANALYSIS
3.1 INTRODUCTION
Analytical instrumentation plays an important role in the production and evaluation of new
products and in the protection of consumers and environment. It is used in checking the quality
of raw materials such as substances used in integrated circuit chips, detection and estimation of
impurities to assure safe foods, drugs, water and air, process optimization and control, quality
check of finished products and research and development. Most of the modern instruments are
microprocessor/computer controlled with user friendly software for collection of data, analysis
and presentation.
This chapter deals with the different types of analytical instrumental methods that find use in a
variety of industries. These include molecular spectroscopic methods, thermal methods of
analysis, X-ray diffraction, scanning electron microscope and sensors.
3.2 SPECTROSCOPY
It is the study of interaction of electromagnetic radiation with matter consisting of atoms and
molecules. When a substance is irradiated with electromagnetic radiation, the energy of the
incident photons may be transferred to atoms and molecules raising their energy from ground
state level to excited state. This process is known as absorption and the resultant spectrum is
known as absorption spectrum. The process of absorption can occur only when the energy
difference between the two levels E is exactly matched by the energy of the incident photons as
given by the equation
E = h? = hc/?
where h is Planck’s constant(6.63 x 10
-34
Js), ? is the frequency of incident radiation, c is the
velocity of light and ? is the wavelength of the incident radiation. The excited state atoms and
molecules then relax to the ground state by spontaneous emission of radiation. The frequency of
the radiation emitted depends on E.
The energy changes that occur in atoms and molecules during interaction with different regions
of electromagnetic radiation are given below.
Radiation
absorbed
Energy of the
radiation
(J/mole)
Effect on the
atoms/molecules
Applications
Introduction, absorption of radiation, UV-Visible Spectrophotometer: Instrumentation and
application, IR Spectrophotometer: Instrumentation and application, Thermal methods of analysis-
TGA, DTA, DSC, Sensors: Oxygen and Glucose sensor, Cyclic Voltammetry for redox system.
?-radiation
> 10
9
Change in nuclear
configuration
Used for cancer radiotherapy.
X- radiation
10
7
- 10
9
Change in core electron
distribution
Chemical crystallography,
qualitative and quantitative
analysis.
Ultraviolet
and Visible
radiation
10
5
-10
7
Change in valence shell
electron distribution.
In qualitative and quantitative
analysis.
Infra red rays
10
3
-10
5
Change in the vibrational
and rotational energy
levels
Detection of functional groups in
compounds, calculation of force
constant, bond length, etc., and in
quantitative analysis
Microwave
radiation
10-10
3
Change in rotational
energy levels
Calculation of force constant,
bond length , bond angle, etc.
Radio
frequency
10
-3
- 10
Changes in nuclear and
electron spin in the
presence of external
magnetic field.
Detection of proton environment
and paramagnetic ions.
3.2.1 UV-Visible spectroscopy
The UV –Visible spectroscopy is also known as electronic absorption spectroscopy as molecules
absorb radiation resulting in transitions between electronic energy levels. Absorption of radiation
in the UV (wavelength range 190-400nm) and visible (wavelength 400–800nm) regions result in
transitions between electronic energy levels. The principle of electronic transitions and the
instruments required to record electronic transitions are common for both the regions. The
electronic transition occurs based on Franck Condon principle which states that electronic
transition takes place so rapidly that a vibrating molecule does not change its inter-nuclear
distance appreciably during the transition.
Polyatomic organic molecules, according to molecular orbital theory, have valence shell
electronic energy structure as shown in Fig 3.1.
Page 3
INSTRUMENTAL METHODS OF ANALYSIS
3.1 INTRODUCTION
Analytical instrumentation plays an important role in the production and evaluation of new
products and in the protection of consumers and environment. It is used in checking the quality
of raw materials such as substances used in integrated circuit chips, detection and estimation of
impurities to assure safe foods, drugs, water and air, process optimization and control, quality
check of finished products and research and development. Most of the modern instruments are
microprocessor/computer controlled with user friendly software for collection of data, analysis
and presentation.
This chapter deals with the different types of analytical instrumental methods that find use in a
variety of industries. These include molecular spectroscopic methods, thermal methods of
analysis, X-ray diffraction, scanning electron microscope and sensors.
3.2 SPECTROSCOPY
It is the study of interaction of electromagnetic radiation with matter consisting of atoms and
molecules. When a substance is irradiated with electromagnetic radiation, the energy of the
incident photons may be transferred to atoms and molecules raising their energy from ground
state level to excited state. This process is known as absorption and the resultant spectrum is
known as absorption spectrum. The process of absorption can occur only when the energy
difference between the two levels E is exactly matched by the energy of the incident photons as
given by the equation
E = h? = hc/?
where h is Planck’s constant(6.63 x 10
-34
Js), ? is the frequency of incident radiation, c is the
velocity of light and ? is the wavelength of the incident radiation. The excited state atoms and
molecules then relax to the ground state by spontaneous emission of radiation. The frequency of
the radiation emitted depends on E.
The energy changes that occur in atoms and molecules during interaction with different regions
of electromagnetic radiation are given below.
Radiation
absorbed
Energy of the
radiation
(J/mole)
Effect on the
atoms/molecules
Applications
Introduction, absorption of radiation, UV-Visible Spectrophotometer: Instrumentation and
application, IR Spectrophotometer: Instrumentation and application, Thermal methods of analysis-
TGA, DTA, DSC, Sensors: Oxygen and Glucose sensor, Cyclic Voltammetry for redox system.
?-radiation
> 10
9
Change in nuclear
configuration
Used for cancer radiotherapy.
X- radiation
10
7
- 10
9
Change in core electron
distribution
Chemical crystallography,
qualitative and quantitative
analysis.
Ultraviolet
and Visible
radiation
10
5
-10
7
Change in valence shell
electron distribution.
In qualitative and quantitative
analysis.
Infra red rays
10
3
-10
5
Change in the vibrational
and rotational energy
levels
Detection of functional groups in
compounds, calculation of force
constant, bond length, etc., and in
quantitative analysis
Microwave
radiation
10-10
3
Change in rotational
energy levels
Calculation of force constant,
bond length , bond angle, etc.
Radio
frequency
10
-3
- 10
Changes in nuclear and
electron spin in the
presence of external
magnetic field.
Detection of proton environment
and paramagnetic ions.
3.2.1 UV-Visible spectroscopy
The UV –Visible spectroscopy is also known as electronic absorption spectroscopy as molecules
absorb radiation resulting in transitions between electronic energy levels. Absorption of radiation
in the UV (wavelength range 190-400nm) and visible (wavelength 400–800nm) regions result in
transitions between electronic energy levels. The principle of electronic transitions and the
instruments required to record electronic transitions are common for both the regions. The
electronic transition occurs based on Franck Condon principle which states that electronic
transition takes place so rapidly that a vibrating molecule does not change its inter-nuclear
distance appreciably during the transition.
Polyatomic organic molecules, according to molecular orbital theory, have valence shell
electronic energy structure as shown in Fig 3.1.
Fig.3.1 Valence shell electronic structure of polyatomic molecules and possible electronic
transitions
In most of the organic molecules, the bonding and non-bonding molecular orbitals are filled, and
the anti-bonding orbitals are vacant. The various electronic transitions that can take place include
(i) s-s
*
(ii) n-s
*
(iii) p-p
*
and (iv) n-p
*
. The relative energy changes involved in these transitions
are in the increasing order n-p
*
< p-p
*
~ n-s
*
<< s-s
*
.
n-p
*
, p-p
*
and
n-s
*
transitions account for the absorption in 200 – 800 nm region of the
electromagnetic spectrum. On the other hand, s-s
*
transition occur in vacuum UV region below
200 nm.
3.2.2 Laws of Absorption
The fraction of the photons absorbed by the molecule at a given frequency depends on
1. The nature of the absorbing molecules
2. The concentration of the molecules (C). The higher the molar concentration, the higher is the
absorption of photons.
3. The length of the path of the radiation through the substance or the thickness of the absorbing
medium. Larger the path length (in cm), larger is the number of molecules exposed and
greater is the probability of photons being absorbed.
Page 4
INSTRUMENTAL METHODS OF ANALYSIS
3.1 INTRODUCTION
Analytical instrumentation plays an important role in the production and evaluation of new
products and in the protection of consumers and environment. It is used in checking the quality
of raw materials such as substances used in integrated circuit chips, detection and estimation of
impurities to assure safe foods, drugs, water and air, process optimization and control, quality
check of finished products and research and development. Most of the modern instruments are
microprocessor/computer controlled with user friendly software for collection of data, analysis
and presentation.
This chapter deals with the different types of analytical instrumental methods that find use in a
variety of industries. These include molecular spectroscopic methods, thermal methods of
analysis, X-ray diffraction, scanning electron microscope and sensors.
3.2 SPECTROSCOPY
It is the study of interaction of electromagnetic radiation with matter consisting of atoms and
molecules. When a substance is irradiated with electromagnetic radiation, the energy of the
incident photons may be transferred to atoms and molecules raising their energy from ground
state level to excited state. This process is known as absorption and the resultant spectrum is
known as absorption spectrum. The process of absorption can occur only when the energy
difference between the two levels E is exactly matched by the energy of the incident photons as
given by the equation
E = h? = hc/?
where h is Planck’s constant(6.63 x 10
-34
Js), ? is the frequency of incident radiation, c is the
velocity of light and ? is the wavelength of the incident radiation. The excited state atoms and
molecules then relax to the ground state by spontaneous emission of radiation. The frequency of
the radiation emitted depends on E.
The energy changes that occur in atoms and molecules during interaction with different regions
of electromagnetic radiation are given below.
Radiation
absorbed
Energy of the
radiation
(J/mole)
Effect on the
atoms/molecules
Applications
Introduction, absorption of radiation, UV-Visible Spectrophotometer: Instrumentation and
application, IR Spectrophotometer: Instrumentation and application, Thermal methods of analysis-
TGA, DTA, DSC, Sensors: Oxygen and Glucose sensor, Cyclic Voltammetry for redox system.
?-radiation
> 10
9
Change in nuclear
configuration
Used for cancer radiotherapy.
X- radiation
10
7
- 10
9
Change in core electron
distribution
Chemical crystallography,
qualitative and quantitative
analysis.
Ultraviolet
and Visible
radiation
10
5
-10
7
Change in valence shell
electron distribution.
In qualitative and quantitative
analysis.
Infra red rays
10
3
-10
5
Change in the vibrational
and rotational energy
levels
Detection of functional groups in
compounds, calculation of force
constant, bond length, etc., and in
quantitative analysis
Microwave
radiation
10-10
3
Change in rotational
energy levels
Calculation of force constant,
bond length , bond angle, etc.
Radio
frequency
10
-3
- 10
Changes in nuclear and
electron spin in the
presence of external
magnetic field.
Detection of proton environment
and paramagnetic ions.
3.2.1 UV-Visible spectroscopy
The UV –Visible spectroscopy is also known as electronic absorption spectroscopy as molecules
absorb radiation resulting in transitions between electronic energy levels. Absorption of radiation
in the UV (wavelength range 190-400nm) and visible (wavelength 400–800nm) regions result in
transitions between electronic energy levels. The principle of electronic transitions and the
instruments required to record electronic transitions are common for both the regions. The
electronic transition occurs based on Franck Condon principle which states that electronic
transition takes place so rapidly that a vibrating molecule does not change its inter-nuclear
distance appreciably during the transition.
Polyatomic organic molecules, according to molecular orbital theory, have valence shell
electronic energy structure as shown in Fig 3.1.
Fig.3.1 Valence shell electronic structure of polyatomic molecules and possible electronic
transitions
In most of the organic molecules, the bonding and non-bonding molecular orbitals are filled, and
the anti-bonding orbitals are vacant. The various electronic transitions that can take place include
(i) s-s
*
(ii) n-s
*
(iii) p-p
*
and (iv) n-p
*
. The relative energy changes involved in these transitions
are in the increasing order n-p
*
< p-p
*
~ n-s
*
<< s-s
*
.
n-p
*
, p-p
*
and
n-s
*
transitions account for the absorption in 200 – 800 nm region of the
electromagnetic spectrum. On the other hand, s-s
*
transition occur in vacuum UV region below
200 nm.
3.2.2 Laws of Absorption
The fraction of the photons absorbed by the molecule at a given frequency depends on
1. The nature of the absorbing molecules
2. The concentration of the molecules (C). The higher the molar concentration, the higher is the
absorption of photons.
3. The length of the path of the radiation through the substance or the thickness of the absorbing
medium. Larger the path length (in cm), larger is the number of molecules exposed and
greater is the probability of photons being absorbed.
Lambert’s law
When a monochromatic beam of radiation passes through an absorbing medium, the intensity of
the transmitted radiation decreases exponentially with the thickness of the absorbing medium.
The law is expressed as
I
t
= I
o
10
–kx
(1)
I
t
and I
o
are the intensities of the transmitted and incident beams of radiations, x is the thickness
of the absorbing medium and k is a constant.
Beer’s law
When a monochromatic beam of radiation passes through an absorbing medium, the intensity of
the transmitted radiation decreases exponentially with the concentration of the absorbing
substance. The law is expressed as
I
t
= I
o
10
– k’C
(2)
where C is the molar concentration of the absorbing substance and k’ is another constant.
Beer-Lambert’s law
When a beam of monochromatic radiation is passed through a transparent absorbing medium, the
decrease in the intensity of radiation is directly proportional to the concentration of the absorbing
substance and the thickness of the absorbing medium.
-dI = kC dx
I
where I is the intensity of radiation, C is the molar concentration of the absorbing species, x is
the thickness of the absorbing medium and k is the proportionality constant. If I
o
is the intensity
of incident radiation and I is the intensity of transmitted radiation, after passing through a path
length (thickness) of l cm in the solution, and upon integrating the above equation, between the
limits I = I
o
when x= 0 and I= I at x= l, we get,
?
?
ln
= - kCl
2.303 log
= -kCl
Page 5
INSTRUMENTAL METHODS OF ANALYSIS
3.1 INTRODUCTION
Analytical instrumentation plays an important role in the production and evaluation of new
products and in the protection of consumers and environment. It is used in checking the quality
of raw materials such as substances used in integrated circuit chips, detection and estimation of
impurities to assure safe foods, drugs, water and air, process optimization and control, quality
check of finished products and research and development. Most of the modern instruments are
microprocessor/computer controlled with user friendly software for collection of data, analysis
and presentation.
This chapter deals with the different types of analytical instrumental methods that find use in a
variety of industries. These include molecular spectroscopic methods, thermal methods of
analysis, X-ray diffraction, scanning electron microscope and sensors.
3.2 SPECTROSCOPY
It is the study of interaction of electromagnetic radiation with matter consisting of atoms and
molecules. When a substance is irradiated with electromagnetic radiation, the energy of the
incident photons may be transferred to atoms and molecules raising their energy from ground
state level to excited state. This process is known as absorption and the resultant spectrum is
known as absorption spectrum. The process of absorption can occur only when the energy
difference between the two levels E is exactly matched by the energy of the incident photons as
given by the equation
E = h? = hc/?
where h is Planck’s constant(6.63 x 10
-34
Js), ? is the frequency of incident radiation, c is the
velocity of light and ? is the wavelength of the incident radiation. The excited state atoms and
molecules then relax to the ground state by spontaneous emission of radiation. The frequency of
the radiation emitted depends on E.
The energy changes that occur in atoms and molecules during interaction with different regions
of electromagnetic radiation are given below.
Radiation
absorbed
Energy of the
radiation
(J/mole)
Effect on the
atoms/molecules
Applications
Introduction, absorption of radiation, UV-Visible Spectrophotometer: Instrumentation and
application, IR Spectrophotometer: Instrumentation and application, Thermal methods of analysis-
TGA, DTA, DSC, Sensors: Oxygen and Glucose sensor, Cyclic Voltammetry for redox system.
?-radiation
> 10
9
Change in nuclear
configuration
Used for cancer radiotherapy.
X- radiation
10
7
- 10
9
Change in core electron
distribution
Chemical crystallography,
qualitative and quantitative
analysis.
Ultraviolet
and Visible
radiation
10
5
-10
7
Change in valence shell
electron distribution.
In qualitative and quantitative
analysis.
Infra red rays
10
3
-10
5
Change in the vibrational
and rotational energy
levels
Detection of functional groups in
compounds, calculation of force
constant, bond length, etc., and in
quantitative analysis
Microwave
radiation
10-10
3
Change in rotational
energy levels
Calculation of force constant,
bond length , bond angle, etc.
Radio
frequency
10
-3
- 10
Changes in nuclear and
electron spin in the
presence of external
magnetic field.
Detection of proton environment
and paramagnetic ions.
3.2.1 UV-Visible spectroscopy
The UV –Visible spectroscopy is also known as electronic absorption spectroscopy as molecules
absorb radiation resulting in transitions between electronic energy levels. Absorption of radiation
in the UV (wavelength range 190-400nm) and visible (wavelength 400–800nm) regions result in
transitions between electronic energy levels. The principle of electronic transitions and the
instruments required to record electronic transitions are common for both the regions. The
electronic transition occurs based on Franck Condon principle which states that electronic
transition takes place so rapidly that a vibrating molecule does not change its inter-nuclear
distance appreciably during the transition.
Polyatomic organic molecules, according to molecular orbital theory, have valence shell
electronic energy structure as shown in Fig 3.1.
Fig.3.1 Valence shell electronic structure of polyatomic molecules and possible electronic
transitions
In most of the organic molecules, the bonding and non-bonding molecular orbitals are filled, and
the anti-bonding orbitals are vacant. The various electronic transitions that can take place include
(i) s-s
*
(ii) n-s
*
(iii) p-p
*
and (iv) n-p
*
. The relative energy changes involved in these transitions
are in the increasing order n-p
*
< p-p
*
~ n-s
*
<< s-s
*
.
n-p
*
, p-p
*
and
n-s
*
transitions account for the absorption in 200 – 800 nm region of the
electromagnetic spectrum. On the other hand, s-s
*
transition occur in vacuum UV region below
200 nm.
3.2.2 Laws of Absorption
The fraction of the photons absorbed by the molecule at a given frequency depends on
1. The nature of the absorbing molecules
2. The concentration of the molecules (C). The higher the molar concentration, the higher is the
absorption of photons.
3. The length of the path of the radiation through the substance or the thickness of the absorbing
medium. Larger the path length (in cm), larger is the number of molecules exposed and
greater is the probability of photons being absorbed.
Lambert’s law
When a monochromatic beam of radiation passes through an absorbing medium, the intensity of
the transmitted radiation decreases exponentially with the thickness of the absorbing medium.
The law is expressed as
I
t
= I
o
10
–kx
(1)
I
t
and I
o
are the intensities of the transmitted and incident beams of radiations, x is the thickness
of the absorbing medium and k is a constant.
Beer’s law
When a monochromatic beam of radiation passes through an absorbing medium, the intensity of
the transmitted radiation decreases exponentially with the concentration of the absorbing
substance. The law is expressed as
I
t
= I
o
10
– k’C
(2)
where C is the molar concentration of the absorbing substance and k’ is another constant.
Beer-Lambert’s law
When a beam of monochromatic radiation is passed through a transparent absorbing medium, the
decrease in the intensity of radiation is directly proportional to the concentration of the absorbing
substance and the thickness of the absorbing medium.
-dI = kC dx
I
where I is the intensity of radiation, C is the molar concentration of the absorbing species, x is
the thickness of the absorbing medium and k is the proportionality constant. If I
o
is the intensity
of incident radiation and I is the intensity of transmitted radiation, after passing through a path
length (thickness) of l cm in the solution, and upon integrating the above equation, between the
limits I = I
o
when x= 0 and I= I at x= l, we get,
?
?
ln
= - kCl
2.303 log
= -kCl
log
=
or
log
= e Cl (where e = k/2.303)
e is the molar absorptivity or molar extinction coefficient, and logI /I
o
= A which is known as
the absorbance of the material.
A = e C l (3)
Thus absorbance A, also known as optical density, is directly proportional to (i) the
concentration C of the absorbing species and (ii) the path length l and has no units. Eq. (3) is the
mathematical expression for Lambert’s Beer law.
e is defined as the absorbance of the solution of unit molar concentration (1M) placed in a cell of
path length one cm. If C is expressed in mol dm
-3
, then the unit for e is dm
3
mol
-1
cm
-1
.
Limitations of Beer –Lambert’s law
Beer-Lambert’s law is strictly valid only in dilute solutions. For dilute solutions, a linear
relationship is exhibited by a plot of absorbance (A) as a function of concentration of the
absorbing substance (C), as shown in Fig 3.2.
Fig 3.2 Plot of Absorbance versus Concentration
(i) Real deviations occur at higher concentration of the absorbing species. At higher
concentrations (>10
-3
M), there is a change in the refractive index of the solution.
(ii) Chemical deviations occur when there is more than one absorbing species present in the
solution. When the absorbing molecules associate or dissociate in the solution, there is a
change in the number of absorbing species.
(iii)Instrumental deviation occurs due to changes in absorptivity of the species as a function of
instrumental bandwidth.
Transmittance T
Read More