EmSAT Achieve Exam  >  EmSAT Achieve Notes  >  Chemistry for EmSAT Achieve  >  Instrumental Techniques

Instrumental Techniques | Chemistry for EmSAT Achieve PDF Download

Table of contents
Instrumental Methods of Analysis
Introduction
Types of Analytical Instrumental Methods
Types of Radiations
Applications of UV-Visible Spectroscopy
Molecular Orbital Theory
Valence Shell Electronic Structure of Polyatomic Molecules
Laws of Absorption
Laws Related to Absorption of Radiation
Lambert's Law
Beer's Law
Beer-Lambert's Law
Key Concepts of Lambert's Beer Law
Definition of &epsilon (Epsilon)
Limitations of Beer-Lambert's Law
Transmittance and UV-Visible Spectrophotometer
Transmittance
Instrumentation of UV-Visible Spectrophotometer
Summary of UV-Visible Spectroscopy
UV-Visible Spectrophotometry Applications Summary
Transitions in UV Light Spectrum
UV-Visible Spectrum of Benzene
Applications of UV-Visible Spectrophotometry
Purity Check
Quantitative Analysis
Determination of Dissociation Constants
Study Areas
IR Spectroscopy
Definition
Principle
Applications of IR Spectroscopy
Identification of Functional Groups and Structure Elucidation
Identification of Substances
Studying the Progress of a Reaction
Detection of Impurities
Base Line Technique
Thermal Analysis
Thermogravimetry (TG) Overview
Introduction to Thermal Analysis
Components of Thermal Analysis Instrumentation
Thermogravimetry (TG) Explained
Instrumentation Details
Summary of Thermal Analysis Process
Thermal Reactions in Calcium Oxalate Decomposition
Thermal Decomposition Processes
CuSO4.5H2O Thermal Decomposition
Thermal Gravimetric Analysis (TGA) Overview
Key Temperature Ranges and Transformations
Applications of Thermal Gravimetric Analysis (TGA)
Differential Thermal Analysis (DTA)
Applications of Thermal Analysis
Sample Holder Assembly
Microprocessor Controlled Furnace
Facility to Control Atmosphere
Data Processor and Recorder
Idealized DTA Curve
Applications of DTA
Differential Temperature
Endothermic
Exothermic
DTA Applications
Examples of DTA
DTA Thermogram
Differential Scanning Calorimetry (DSC)
Definition of DSC:
Principle of DSC:
Instrumentation of DSC:
Block Diagram of DSC Instrument:
Applications of Differential Scanning Calorimetry (DSC)
Sensors
Oxygen Sensors
Figure 3.15 Clarke Electrochemical O2 Sensor
Limitations
Paramagnetic Oxygen Sensor
Zirconia Oxygen Sensors
Porous Pt Oxygen Sensor and Applications
Zirconia-Based Oxygen Sensor
Applications of Oxygen Sensors
Glucose Sensors
Cyclic Voltammetry
Example Graph:
Introduction to Cyclic Voltammetry
Application
Questions - Part A
Paraphrased Explanation
Application of Cyclic Voltammetry
Test Questions - Part A
Key Concepts in Analytical Chemistry
Different Types of Oxygen Sensors
Distinguishing DSC and DTA
Understanding Glucose Sensors
Defining Cyclic Voltammetry
Advantages of Cyclic Voltammetry
Applications of Cyclic Voltammetry
Instrumentation and Principles of Analytical Techniques
Laws of Absorption and Beer-Lambert's Law
UV-Visible Spectrophotometer Instrumentation
IR Spectrophotometer Instrumentation
Thermogravimetry Principles and Applications
Differential Thermal Analysis Fundamentals
Differential Scanning Calorimetry Overview
Understanding Sensors and Electrochemical Techniques
Overview of Sensors
Types of Oxygen Sensors
Exploring Biosensors and Glucose Sensors
Working Principle of Cyclic Voltammetry
Educational Content Creation
Content Creation Specialization
Content Formatting Expertise
Test Question Creation

Instrumental Methods of Analysis

  • Introduction

    Analytical instrumentation is crucial for creating and assessing new products, ensuring consumer safety, and environmental protection.

    It is used in various applications such as quality control of raw materials, impurity detection in food and drugs, process optimization, and research.

    Modern instruments are typically computer-controlled with user-friendly software for data collection and analysis.

  • Types of Analytical Instrumental Methods

    • Spectroscopy

      Spectroscopy involves studying how electromagnetic radiation interacts with atoms and molecules.

      When irradiated, atoms and molecules absorb energy, moving from a ground state to an excited state, creating an absorption spectrum.

      This absorption process matches the energy difference between levels with the incident photons' energy.

      For example, UV-Vis Spectrophotometer and IR Spectrophotometer are used to analyze the absorption of radiation by substances.

    • Thermal Methods of Analysis

      Thermal methods like TGA (Thermogravimetric Analysis), DTA (Differential Thermal Analysis), and DSC (Differential Scanning Calorimetry) are employed to study materials' thermal properties.

    • Sensors

      Sensors like Oxygen and Glucose sensors play a vital role in detecting and measuring specific substances in various applications.

    • Cyclic Voltammetry

      Cyclic Voltammetry is utilized for studying redox reactions by measuring current as a function of voltage.

Types of Radiations

  • y-radiation: Greater than 10 degrees. Involves a change in nuclear configuration. Utilized in cancer radiotherapy.
  • X-radiation: Ranges from 10^7 to 10^9. Involves a change in core electron distribution and valence shell electron distribution.
  • Ultraviolet and Visible radiation: Ranges from 10^5 to 10^7.
  • Infrared rays: Ranges from 10^3 to 10^5. Involves changes in vibrational and rotational energy levels.
  • Microwave: Ranges from 10 to 10^3. Involves changes in rotational radiation energy levels.
  • Radio: Ranges from 10^-3 to 10. Involves changes in nuclear and frequency and electron spin in the presence of an external magnetic field. Used in chemical crystallography, qualitative, and quantitative analysis.

Applications of UV-Visible Spectroscopy

  • UV-Visible spectroscopy, also known as electronic absorption spectroscopy, involves molecules absorbing radiation leading to transitions between electronic energy levels.
  • Electronic transitions in the UV (190-400nm) and visible (400-800nm) regions result in transitions between electronic energy levels.
  • The Franck-Condon principle explains that electronic transitions occur so rapidly that a vibrating molecule maintains its inter-nuclear distance during the transition.

Molecular Orbital Theory

  • Polyatomic organic molecules, as per molecular orbital theory, have a valence shell electronic energy structure.
  • Energy levels include antibonding, non-bonding, and bonding orbitals.

Valence Shell Electronic Structure of Polyatomic Molecules

  • In organic molecules, bonding and non-bonding molecular orbitals are filled, while anti-bonding orbitals remain vacant.
  • Electronic transitions can occur, including:
    • σ-σ*
    • n-σ*
    • λ-λ
    • n-π*
  • The energy changes in these transitions follow the order: n-λˆ<λ-ï*~>< />
  • n-ê, λ-ë³, and n-σ* transitions lead to absorption in the 200 - 800 nm region of the electromagnetic spectrum. On the contrary, σ-σ* transitions occur in the vacuum UV region below 200 nm.

Laws of Absorption

  • The fraction of photons absorbed by a molecule at a specific frequency depends on:
    • The nature of the absorbing molecules.
    • The concentration of the molecules (C). Higher molar concentration results in increased photon absorption.
    • The length of the radiation path through the substance or the thickness of the absorbing medium. A greater path length (in cm) exposes more molecules, increasing the probability of photon absorption.
  • Lambert's Law

    When a monochromatic radiation beam passes through an absorbing medium, the intensity of the transmitted radiation decreases exponentially with the medium's thickness.

    The law is represented as: I₁ = I₀e^(-kx)

    Here, I₁ and I₀ are the intensities of transmitted and incident radiation beams respectively, x is the thickness of the medium, and k is a constant.

  • Beer's Law

    Describes how the intensity of transmitted radiation decreases exponentially with the concentration of the absorbing substance when a monochromatic beam passes through it.

    The law can be expressed as: I₁ = I₀e^(-k'C)

    Where C represents the molar concentration of the absorbing substance and k' is another constant.

  • Beer-Lambert's Law

    States that the intensity of radiation passing through a transparent absorbing medium is directly proportional to the concentration of the substance and the medium's thickness.

    The equation is: -dI = kC dx

    Here, I is the intensity of radiation, C is the molar concentration, x is the thickness of the medium, and k is the proportionality constant.

    Upon integration, the equation simplifies to: log(I₀/I) = -kCl

Key Concepts of Lambert's Beer Law

  • The absorbance (A) of a material is directly proportional to the concentration (C) of the absorbing species and the path length (l).
  • The mathematical expression for this relationship is represented by A = &epsilon * C * l, where &epsilon is the molar absorptivity or molar extinction coefficient.
  • Absorbance, also known as optical density, has no units and is a measure of how much light is absorbed by a substance.

Definition of &epsilon (Epsilon)

  • &epsilon is defined as the absorbance of a solution of unit molar concentration (1M) placed in a cell of path length one cm.
  • When the concentration (C) is expressed in mol dm^3, the unit for &epsilon is dm^3 mol^-1 cm^-1.

Limitations of Beer-Lambert's Law

  • Beer-Lambert's law is strictly valid only in dilute solutions where a linear relationship is observed between absorbance (A) and concentration (C).
  • Real deviations occur at higher concentrations due to changes in the refractive index of the solution.
  • Chemical deviations may happen when multiple absorbing species are present, leading to changes in the number of absorbing molecules.
  • Instrumental deviations can occur due to variations in the absorptivity of the species with respect to instrumental bandwidth.

Transmittance and UV-Visible Spectrophotometer

Transmittance

  • Transmittance refers to the portion of light that passes through a substance.
  • It is calculated as T = I / Io, where I is the intensity of transmitted light and Io is the intensity of incident light.
  • The relationship between transmittance (T) and absorbance (A) is given by A = -log T.
  • Transmittance T is often expressed as a percentage.

Instrumentation of UV-Visible Spectrophotometer

  • The UV-Visible spectrophotometer is an instrument used to analyze the spectra of molecules.
  • It consists of several key components:
    • Source of radiation
    • Monochromator
    • Sample cell
    • Detector
    • Display/Recorder
  • A typical block diagram of a UV-Visible spectrophotometer includes:
    • Radiation source: Uses hydrogen discharge lamp for UV and tungsten filament lamp for visible light.
    • Monochromator: Selects a narrow range of wavelengths for analysis.
    • Sample cell: Holds the substance being analyzed.
    • Detector: Measures the intensity of transmitted light.
    • Display/Recorder: Shows or records the data obtained from the analysis.

Summary of UV-Visible Spectroscopy

  • Monochromator:

    It separates polychromatic radiation from a source into a narrow range of wavelengths. Quartz prisms or gratings are commonly used for UV and visible light. Examples include the 60° Cornu quartz prism and 30° Littrow prisms for UV, visible light can use a glass prism.

  • Sample Holder (Cells or Cuvettes):

    Containers for samples must be transparent to UV and visible radiation. Quartz cuvettes are used for both UV and visible regions, while glass cuvettes are suitable for visible light. The standard path length for these cuvettes is typically 1 cm.

  • Sector Mirror:

    This component splits the monochromatic radiation into two parallel beams that pass through sample and reference cells before reaching the detector.

  • Solvents for UV Region:

    Solvents like water, methyl alcohol, ethyl alcohol, chloroform, and hexane are used for solutions in the UV and visible regions. 95% ethyl alcohol is commonly used in the UV region due to its properties.

  • Detectors:

    Photovoltaic cells, photoemissive cells, or photomultiplier tubes are utilized to convert incident photons into electric current for detection.

  • Display/Recorder:

    The recorder and display unit synchronize to record detector signals as transmittance or absorbance units against the wavelength of incident radiation.

  • UV-Visible Spectrometer Operation:

    A beam of light is split into sample and reference beams in the spectrometer. By comparing their intensities at various wavelengths, absorption data is collected to generate an absorption spectrum.

  • Characteristics of UV and Visible Spectra:

    • Wavelength of Absorption Maximum (λmax):

      This is the wavelength where maximum absorption occurs, varying for different molecules.

    • Molar Absorptivity (ε):

      Relates to the height of the absorption band for a given compound concentration. Important for characterizing and quantifying compounds.

    • Chromophores:

      Groups like C=C, C=N, N=N, C=O are responsible for absorption due to their structure. They are key in characterizing compounds.

UV-Visible Spectrophotometry Applications Summary

Transitions in UV Light Spectrum

  • Auxochromes, like -OH, -Cl, -OR, NR2, etc., alter chromophore absorption in UV light.

UV-Visible Spectrum of Benzene

  • Illustrative diagram showing the electronic absorption spectrum of benzene in hexane.

Applications of UV-Visible Spectrophotometry

  • Qualitative Analysis
    • UV-Visible spectra help identify unknown organic samples.
    • Key Observations
      • Absorption patterns below 200 nm.
      • Strong absorption between 200 and 250 nm.
      • Weak absorption near 300 nm.
    • Possible Conclusions
      • Molecules with only σ bonds, lone pairs, or isolated double bonds (e.g., CH2=CH2 at 180nm).
      • Conjugated double bonds show increased absorption maxima (e.g., butadiene at 210nm).
      • Long chain conjugated molecules like polyenes and carotenes absorb in the visible region.
      • Presence of aromatic rings indicated by absorption at specific wavelengths (e.g., benzene at 250nm).
      • Presence of carbonyl compounds (C=O).

Purity Check

  • Value comparison used to identify substances.
  • & value affected by absorbing substance's nature and incident light wavelength (λ).
  • Sample purity checked by comparing & values with standard sample.
  • Deviations indicate impurities or adulteration.

Quantitative Analysis

  • Measurement of organic compounds and inorganic complexes using Lambert's Beer law.
  • A = & Cl formula for absorbance measurement.
  • Linear plot of Absorbance (A) vs. Concentration (C).

Determination of Dissociation Constants

  • Weak acids/bases dissociation constants from absorption spectra changes with pH.

Study Areas

  • Investigation of chemical reaction kinetics.
  • Examination of molecules' electronic structure e.g., vitamins, steric hindrance detection.

IR Spectroscopy

Definition

  • Study of infrared region (700nm to 1000μm) in the electromagnetic spectrum.
  • Transitions between vibrational levels.

Principle

  • IR spectra from molecule energy absorption in infrared region.
  • Transitions occur between vibrational levels.

Regions

  • Near IR: 12500 to 4000 cm⁻¹
  • IR: 4000 to 670 cm⁻¹
  • Far IR: 670 to 50 cm⁻¹

Theory of IR Absorption

  • IR induces vibrational and rotational changes, not electronic transitions.
  • Molecule absorbs IR if net dipole moment changes during vibration.
  • Vibrational Modes in Molecules:
    • Overview:

      The energy from infrared (IR) radiation should match the energy gap between two vibrational levels in a molecule.

    • Vibrational Movements:
      • Types of Vibrations:
        • Stretching Vibration:
          • Symmetrical Stretching: Both bonds lengthen or shorten simultaneously.
          • Asymmetrical Stretching: One bond lengthens while the other shortens.
        • Bending Vibration:
          • In-plane Bending:
            • Scissoring: Atoms move towards each other in the same plane.
            • Rocking: Atoms move in the same direction, affecting bond angles.
          • Out-plane Bending:
            • Wagging: Atoms move up and down on one side of the plane.
            • Twisting: One atom moves above while another moves below the plane.
  • Instrumentation
    • The essential components of an infrared spectrophotometer include:
      • A source: Common sources are the Nernst glower and globar, both requiring heating to emit infrared radiation.
      • Sample cells and Sampling techniques: Cells made of Nacl for transparency to IR light. Different techniques for gaseous, liquid, and solid samples.
      • Solvents: Transparent solvents like CCl4, CS2 that dissolve samples completely.
      • Monochromator: Separates radiation into individual wavelengths using materials like NaCl, LiF, CaF2.
      • Detectors: Convert light signals to electrical signals using various types of detectors like photovoltaic cells, bolometers.
      • Amplifier / Recorder: Amplifies and converts electrical signals to percentage transmittance, recording the data.

Illustration of a block diagram of an IR spectrophotometer:

Instrumental Techniques | Chemistry for EmSAT Achieve

Explanation:

The infrared spectrophotometer comprises several crucial components that work together to analyze samples. To begin with, the source of the instrument, such as the Nernst glower or globar, must be heated significantly to emit infrared radiation. Sample handling involves different techniques for gaseous, liquid, and solid samples, each requiring specific sample cells and preparation methods. Solvents play a vital role in ensuring the sample dissolves completely and is transparent to IR light.

The monochromator is responsible for separating radiation into distinct wavelengths, utilizing materials like NaCl, LiF, or CaF2. Detectors then convert the received light signals into electrical signals through various mechanisms like photovoltaic cells or bolometers. Subsequently, the amplifier and recorder process these signals, converting them into percentage transmittance data that is recorded for analysis.

Applications of IR Spectroscopy

  • Identification of Functional Groups and Structure Elucidation

    • IR spectroscopy divides the IR region into group frequency and fingerprint regions.
    • In the group frequency region (4000-1500 cm^-1), peaks corresponding to different functional groups are observed, aiding in functional group determination.
    • The fingerprint region (1500-400 cm^-1) helps identify molecules by characteristic peaks.
  • Identification of Substances

    • IR spectra are utilized to determine if two organic substances are identical based on their absorption bands.
    • Distinct IR spectra imply different compounds, while identical spectra indicate the same substance.
    • Enantiomeric compounds exhibit identical IR spectra, making them indistinguishable via IR spectroscopy.
  • Studying the Progress of a Reaction

    • IR spectroscopy aids in monitoring reaction progress by observing absorption band changes over time.
    • Changes in reactant and product absorption bands indicate reaction advancement.
  • Detection of Impurities

    • Comparing the IR spectrum of a sample with a standard compound helps identify impurities through additional peaks in the spectrum.
  • Quantitative Analysis

    • IR spectroscopy enables the determination of substance quantities, whether in pure form or as a mixture.
    • Quantities are assessed by selecting characteristic peaks corresponding to the substance.

Base Line Technique

  • Comparison of log Io/I₁ of peaks for standard and test samples to determine substance quantity.
  • Limitations:
    • Molecular weight prediction is not possible.
    • Frequent non-adherence to Beer's law in complex spectra.
    • IR spectroscopy does not reveal the relative position of different functional groups in a molecule.

Thermal Analysis

  • Thermal analysis involves monitoring physical property changes like weight, temperature, or enthalpy of a sample material concerning temperature variations.
  • Techniques:
    • Thermogravimetric analysis (TGA): Records sample weight changes with temperature.
    • Differential thermal analysis (DTA): Measures temperature differences between a sample and reference material under controlled heating.
    • Differential scanning calorimetry (DSC): Measures energy differences between a sample and reference material under controlled heating.
  • Applications:
    • Thermal stability and compositional analysis of alloys, mixtures, and corrosion studies.
    • Generation of phase diagrams, study of phase transitions, thermal stability, and polymer characterization.
    • Reaction kinetics and purity analysis of drugs.
  • Thermal events during analysis may include:
    • Phase transitions
    • Melting
    • Sublimation/volatilization
    • Decomposition
    • Glass transition in polymers
    • Oxidation/reduction, etc.

Thermogravimetry (TG) Overview

Introduction to Thermal Analysis

  • Controlled Temperature Programme: Thermal analysis involves subjecting materials to a controlled temperature programme.

Components of Thermal Analysis Instrumentation

  • Furnace: The furnace, temperature sensor, and controlled atmosphere are essential components.
  • Sample and Container: This includes the sample and its container.
  • Sensors: Instruments used for measuring temperature and sample properties.
  • Computer and Equipment: The setup includes data collection, processing equipment, and a display for results.

Thermogravimetry (TG) Explained

  • Definition of Thermogravimetry: TG is a technique that tracks sample weight changes concerning temperature.
  • Principle of TG: Weight changes are monitored as the sample is heated at a controlled rate.
  • Instrumentation: TG's key component is the thermobalance or thermogravimetric analyzer.

Instrumentation Details

  • Sample Handling: Solid samples are placed in a platinum crucible and connected to a microbalance.
  • Thermobalance Function: The balance within the furnace is crucial for accurate measurements.
  • Null Point Balance: A mechanism that detects weight changes in the sample.

Fig 3.5: Block Diagram of TG Apparatus

  • Furnace
  • Control
  • Balance
  • Crucible
  • Sample
  • Data Processor
  • Recorder

Summary of Thermal Analysis Process

  • Initiating Balance Force:
    • A force is activated to return the system to equilibrium. This force is directly proportional to the change in weight experienced.
  • Control of Furnace Atmosphere:
    • The atmosphere within the furnace is regulated by the introduction of gases. Inert gases like nitrogen, helium, or argon, as well as reactive gases such as oxygen and hydrogen, can be utilized for this purpose.
  • Data Processing and Recording:
    • The balance mechanism initially gauges the weight of the sample and continuously tracks weight fluctuations as heat is administered within the furnace.
    • Experimental data from the furnace and balance are gathered and transmitted to a computer for analysis. The computer then records the Thermal Gravimetric (TG) curve.

Thermal Reactions in Calcium Oxalate Decomposition

  • Thermogram and Thermal Reactions:
    • A thermogram for calcium oxalate monohydrate is displayed in Figure 3.6.
    • Different thermal reactions during the heating process of calcium oxalate monohydrate, ranging from 30°C to around 1000°C, are outlined in Table 3.1.
    • The TG curve illustrates distinct regions: plateaus represent stability with no weight loss, while downward steps indicate areas of weight reduction.
  • Summary of Decomposition Stages:
    • 30-130°C: Initial plateau where CaC2O4.H2O remains thermally stable with no mass change.
    • 130-190°C: First downward step involving CaC2O4.H2O converting to CaC2O4 with water loss and decrease in mass.

Thermal Decomposition Processes

  • 190 - 400°C
    • 2nd plateau region: Anhydrous CaC2O4 is stable under heat.
    • No change in mass observed.
    • Mass decrease occurs in this temperature range.
  • 400 - 470°C
    • 2nd downward step: CaC2O4 transforms into CaCO3 and CO due to CO loss.
  • 470 - 700°C
    • 3rd plateau: CaCO3 remains stable under heat.
    • No change in mass is detected.
    • Mass decreases in this temperature range.
  • 700 - 840°C
    • 3rd downward step: CaCO3 converts to CaO and CO2 through CO2 loss.
  • 840 - 1000°C
    • 4th plateau: CaO is thermally stable.
    • No change in mass is observed. The resulting residue is CaO.
  • Thermogram of CuSO4.5H2O
    • Shown in Figure 3.7
    • Thermal events summarized in Table 3.2
  • CuSO4.5H2O Thermal Decomposition

  • Loss of 4H2O:
    • Mass loss due to water evaporation
  • Loss of H2O:
    • Transformation to CuSO4 and H2O, with 3 H2O loss
    • Change in mass observed
  • Loss of SO2 1/2O2:
    • Conversion to CuO or Cu2O
  • Temperature Ranges and Reactions
    • 30 - 90°C
      • 1st plateau: CuSO4.5H2O remains stable
    • 90 - 150°C
      • 1st downward step: CuSO4.5H2O transforms into CuSO4, H2O, and loses 3 H2O
      • Change in mass occurs due to water of crystallization loss
  • Thermal Gravimetric Analysis (TGA) Overview

    • The process involves heating a substance to observe changes in its mass as a function of temperature.
    • Crucial temperature points and reactions are noted during the analysis.

    Key Temperature Ranges and Transformations

    • 150-200°C: Second plateau region where CuSO4 · H2O exhibits thermal stability.
    • 200-275°C: Second downward step involving the transformation of CuSO4 · H2O into CuSO4 and H2O.
    • 275-700°C: Third plateau indicating the stability of anhydrous CuSO4.
    • 700-900°C: Third downward step leads to the formation of CuO from CuSO4, along with the release of SO2.
    • 900-1000°C: Fourth plateau where CuO remains thermally stable.
    • 1000-1100°C: Fourth downward step involves the conversion of 2CuO into Cu2O and O2.
    • No Change in Mass: Indicates the absence of mass alteration.
    • Decrease in Mass due to Loss of H2O: Demonstrates mass reduction resulting from water loss.
    • Decrease in Mass due to Decomposition: Shows mass decrease due to substance breakdown.
    • Reduction of CuO: Conversion of CuO to Cu2O with a decrease in mass occurs.

    Applications of Thermal Gravimetric Analysis (TGA)

    • Analysis of Inorganic Salts and Complexes: TGA is utilized in studying thermal decomposition of materials like catalysts, semiconductors, and fine chemicals.
    • Investigating Polymer Decomposition: TGA helps determine the decomposition temperature of plastics and rubber, aiding in identifying different polymers based on their thermograms.
    • Thermal Stability of Polymers: Different polymers exhibit unique thermograms, with polymers like PTFE, LDPE, PMMA, and PVC showing varying thermal stabilities. PVC notably displays a two-stage decomposition process.
    In this format, the complex information about Thermal Gravimetric Analysis (TGA) has been summarized into a more digestible and organized form using HTML elements. The details are broken down into key sections and points, making it easier to understand and remember.

    Differential Thermal Analysis (DTA)

    • Definition: Differential thermal analysis is a method that measures the temperature difference (ΔT) between a sample and an inert reference material as the sample is uniformly heated.
    • Principle: This technique detects energy absorption or release by the sample during transitions like melting or decomposition.
      • When a sample undergoes a change, its temperature deviates from the reference's temperature, resulting in ΔT.
      • If no change occurs, both the sample and reference remain at the same temperature, resulting in ΔT being zero.
      • Endothermic or exothermic reactions cause temperature variations between the sample and reference, reflected in the DTA thermogram.
    • Instrumentation: The DTA instrument includes a microprocessor-controlled furnace, data processor, recorder, and atmosphere control feature.
      • The instrument's schematic, as shown in Fig 3.9, displays components such as furnace, balance, and data recorder.

    Applications of Thermal Analysis

    • TGA Applications: Thermal gravimetric analysis is valuable in various fields:
      • Studying thermal reactions in pharmaceuticals, coal, minerals, and alloys.
      • Qualitative compound analysis and oxidation studies of alloys.
      • Determining compositions of mixtures, such as distinguishing between calcium and strontium carbonates based on their decomposition temperatures.

    Sample Holder Assembly

    • Solid sample and reference material (often an inert substance like alumina of 10 mg) are placed in a platinum crucible (sample container).
    • The assembly is connected to a sensitive microbalance for measurement.
    • Temperature of the sample and reference is monitored using individual thermocouples.

    Microprocessor Controlled Furnace

    • The entire sample holder assembly is positioned inside the furnace.
    • The sample and reference materials are heated simultaneously at a consistent rate up to 1500°C.
    • A temperature programmer or furnace control maintains a steady heating rate of 1°C/min to 100°C/min.

    Facility to Control Atmosphere

    • The sample and reference chamber allow the circulation of inert gases like nitrogen or reactive gases such as oxygen or air.

    Data Processor and Recorder

    • The temperature difference (ΔT) between the sample and reference thermocouples (S and R) is continually monitored.
    • After amplification, the signal difference is recorded on the y-axis.
    • The furnace temperature is tracked by an independent thermocouple and recorded on the x-axis.
    • Balance and furnace data are collected, sent to a PC for analysis, and a DTA plot of (ΔT) vs. T is generated.

    Idealized DTA Curve

    • DTA curves display peaks that indicate various reactions—endothermic peaks suggest changes in crystallinity or dehydration, while exothermic peaks result from reactions like oxidation.
    • An example curve is shown in Fig 3.10, depicting exothermic and endothermic peaks.

    Applications of DTA

    • Qualitative analysis: DTA measurements offer a swift method to identify minerals, clays, and polymeric materials.
    • For example, Fig 3.11 illustrates the DTA thermogram of calcium oxalate monohydrate in flowing air (O2) undergoing temperature increase at 8°C/min, showing specific peaks for decomposition and oxidation reactions.

    Differential Thermal Analysis (DTA)

    Differential Temperature

    • DTA measures temperature differences

    Endothermic

    • Processes or reactions that absorb heat

    Exothermic

    • Processes or reactions that release heat
    • Example: CaC₂O → CaCO + CO
    • Example: CaC₂O.H₂O → CaC₂O + H₂O
    • Example: CaCO₃ → CaO + CO₂

    DTA Applications

    • Provides insights into fusion, dehydration, oxidation, reduction, adsorption, and solid-state reactions
    • Determines melting and boiling points for organic compounds accurately
    • Utilized in polymer studies, characterization, and qualitative analysis of polymer mixtures
    • Helps in generating phase diagrams and studying phase transitions

    Examples of DTA

    • Illustration of transitions during heating of a polymer
    • Processes like crystallization, oxidation, glass transition, melting, onset of decomposition, etc.

    DTA Thermogram

    • Visual representation of heat flow during processes
    • Example: DTA thermogram of a polymer

    Differential Scanning Calorimetry (DSC)

    • Definition of DSC:

      Differential Scanning Calorimetry (DSC) is a thermal technique that measures the difference in heat flow between a sample and a reference material as the sample's temperature changes linearly.

    • Principle of DSC:

      DSC works by comparing the heat flow needed to maintain the sample and reference materials at the same temperature. When the sample undergoes a reaction, the temperature of the sample changes compared to the reference. The energy change or enthalpy change in the sample is calculated based on this temperature difference.

    • Instrumentation of DSC:

      DSC instruments are similar to Differential Thermal Analysis (DTA) instruments. The setup includes a furnace controlled by a microprocessor, individual heaters and thermocouples for the sample and reference, a data processor, recorder, and controls for the atmosphere.

      • Sample Holder Assembly:

        A sample and a reference material are placed in separate crucibles and heated individually. The temperature difference between them is monitored by thermocouples.

      • Microprocessor Controlled Furnace:

        The entire sample holder assembly is positioned within the furnace. Both the sample and reference are heated at the same rate, and their temperatures are recorded by thermocouples.

    • Block Diagram of DSC Instrument:

      The DSC instrument consists of components such as a furnace, control sensor, heater, data processor, and recorder. The furnace is equipped with heaters and thermocouples for precise temperature measurements.

    Differential Scanning Calorimetry (DSC)

    • Temperature Control

      DSC involves heating samples in a controlled manner up to 700°C. A temperature programmer maintains a constant heating rate, typically between 1°C/min to 100°C/min.

    • Atmosphere Control

      The system allows for the circulation of inert gases like nitrogen or reactive gases such as oxygen or air within the sample and reference chambers.

    • Data Processing and Recording

      The heat flow difference between the sample and reference thermocouples is continuously measured and recorded. This data, along with furnace temperature, is sent to a PC for analysis. A DSC thermogram is generated based on the recorded data.

    • Interpreting DSC Results

      A DSC thermogram provides valuable information about exothermic and endothermic processes. For example, the decomposition of calcium oxalate monohydrate can be visualized through a DSC thermogram.

    Applications of Differential Scanning Calorimetry (DSC)

    • Phase Transitions Study

      DSC is crucial for studying phase transitions and temperature-induced changes in materials like polymers, biological samples, and pharmaceuticals that occur at low temperatures.

    • Enthalpy Calculations

      It is widely used to calculate enthalpy changes such as enthalpy of melting, crystallization, and fusion in polymeric materials.

    • Organic Compounds Analysis

      DSC helps in determining melting, boiling, and decomposition points accurately for organic compounds, aiding in their characterization.

    • Purity Determination

      One of the key applications includes assessing the purity of drug samples, ensuring the quality and integrity of pharmaceutical products.

    Sensors

    • Devices utilized for detecting alterations in physical and chemical quantities are referred to as sensors.

    Oxygen Sensors

    • An oxygen sensor, also known as a lambda sensor, is an electronic tool that quantifies the proportion of oxygen (O2) in the gas or liquid under examination.

    Clarke Electrochemical Oxygen Sensor

    • An electrochemical oxygen analyzer operates based on the electrochemical reduction of O2 at a negatively polarized electrode, commonly known as the "Clark Type" after its creator, Dr. Leland Clark.
    • The Clark electrode comprises a silver anode and a platinum cathode in contact with a 0.01M KCl electrolyte solution.
    • A semi-permeable membrane, typically made of polypropylene or Teflon, covers the electrode's tip, allowing gases to pass while blocking the electrolyte.
    • When the sensor is immersed in a sample solution with dissolved oxygen, O2 diffuses through the membrane and reaches the cathode.
    • The silver/silver chloride (Ag/AgCl) anode supplies electrons for the cathode reaction, with silver at the anode undergoing oxidation.
    • A potential difference of 1.5V is maintained between the two electrodes, polarizing the cathode sufficiently (-0.6V) to reduce all oxygen present.
    • The current flowing through the circuit is directly proportional to the amount of gas crossing the membrane, enabling the measurement of oxygen concentration in the solution.

    Electrode Reactions:

    • Ag anode: 4Ag → 4AgCl + 4e-
    • Pt cathode: O2 + 4H+ + 4e- → 2H2O

    Key Components:

    • Teflon membrane
    • Ag electrode
    • Pt working electrode
    • -0.1 M KCl
    • Sample solution

    Figure 3.15 Clarke Electrochemical O2 Sensor

    Limitations

    • One limitation is the susceptibility of the system to damage when exposed to acidic gases like HCl, H₂S, and SO₂.

    Paramagnetic Oxygen Sensor

    • Oxygen, nitric oxide, and nitrogen dioxide possess paramagnetic properties, being attracted to a magnetic field, with oxygen exhibiting higher paramagnetism.
    • In gaseous oxygen analyzers, a mechanism involving nitrogen-filled glass spheres is utilized to detect oxygen concentration based on magnetic susceptibility.
    • When oxygen molecules flow into the system, they are drawn towards a stronger magnetic field, causing a deviation in the balance of the spheres.
    • The resulting deviation is detected using light-based elements and a feedback loop to maintain balance, with the current in the loop being proportional to oxygen concentration.

    Zirconia Oxygen Sensors

    • Solid-state potentiometric oxygen sensors employ materials like yttria-stabilized zirconia (YSZ) as an electrolyte due to its ionic nature and stability in harsh conditions.
    • By stabilizing zirconia with around 6-12 mol% of yttria, a favorable phase for ionic conduction is achieved at high temperatures.
    • This sensor configuration typically consists of a cell with YSZ acting as a solid electrolyte, enabling high-temperature electrochemical sensing.

    Porous Pt Oxygen Sensor and Applications

    Zirconia-Based Oxygen Sensor

    • A thimble type YSZ-based oxygen sensor comprises a ceramic tube coated with porous platinum on both surfaces to function as anode and cathode electrodes.
    • At temperatures above 450°C, oxygen ions can move through the crystal lattice due to openings, creating a concentration gradient based on partial pressures.
    • The Nernst equation is used to determine the amount of oxygen in the sample gas based on the concentration gradient.
    • The potential difference across the cell at high temperatures is calculated using the Nernst equation.

    Nernst Equation

    • The potential difference (E) is given by: E = 4F(ln(P1/P2))/RT
    • Where:
      • E: potential difference (volts)
      • R: gas constant (8.314 J mol¯¹ K¯¹)
      • T: absolute temperature (K)
      • F: Faraday constant (96500C)
      • P1 & P2: partial pressures of oxygen on either side of the zirconia tube

    Applications of Oxygen Sensors

    • To measure oxygen concentration in exhaust gases of IC engines in vehicles.
    • To determine oxygen partial pressure in breathing gas for deep-sea divers.
    • Utilized in anesthesia monitors, respirators, and oxygen concentrators.

    Glucose Sensors

    • Glucose sensors play a crucial role not only in blood sugar monitoring but also in various industries like food, bio-processing, and fuel cell development.
    • The first amperometric enzyme glucose sensor was created in 1973, focusing on the analysis of hydrogen peroxide production.
    • The enzyme glucose 1-oxidase (GOx) is a key component in enzymatic biosensors due to its high selectivity, sensitivity, and stability.
    • Flavin adenine dinucleotide (FAD) in the redox center of GOx is essential for glucose oxidation, leading to the production of glucolactone.
    • Glucose concentration is determined by measuring the hydrogen peroxide generated, which is oxidized at a platinum anode.
    • The process offers a simple, precise, and selective method for blood glucose testing.

    Cyclic Voltammetry

    • Definition: Cyclic voltammetry (CV) is an electrochemical technique that assesses current in an electrochemical cell under specific voltage conditions.
    • Explanation: In a CV experiment, the working electrode's potential is varied linearly over time, returning to the initial potential after reaching a set value.
    • This cyclic process allows for multiple repetitions and the plotting of current against applied voltage, providing valuable electrochemical information.

    Cyclic Voltammetry

    • Definition of Cyclic Voltammetry: Cyclic voltammetry is a technique commonly employed to examine the electrochemical properties of a substance in a solution by plotting current against applied potential over time.
    • Basic Principle: During a cyclic voltammetry experiment, the electrode potential varies linearly in cyclical phases. The rate of change of voltage over time in each phase is termed the scan rate (measured in V/s).
    • Data Collection: The potential is applied between the working electrode and the reference electrode, while the current is measured between the working electrode and the counter electrode.
    • Plotting Data: The collected data is depicted as a graph of current (i) against applied potential (E). This graph helps in understanding the behavior of the analyte in the system.
    • Interpreting the Graph: During the forward scan, a reducing potential is applied, leading to an increase in cathodic current until the reduction potential of the analyte is reached. The reverse scan shows the re-oxidation of the reduced analyte.
      • Reversibility of Redox Couple: If the redox couple is reversible, the oxidation peak mirrors the reduction peak, providing insights into redox potentials and reaction rates.
      • Cottrell Equation: The relationship between peak current and scan rate, as described by the Cottrell equation, helps understand the electron transfer dynamics and analyte diffusion at the electrode surface.
    • Information from CV Data: Cyclic voltammetry data can offer valuable information about electrochemical reactions, redox potentials, and reaction kinetics based on the shape and characteristics of the peaks observed.

    Example Graph:

    Instrumental Techniques | Chemistry for EmSAT Achieve

    Introduction to Cyclic Voltammetry

    • Cyclic voltammetry is a method commonly used to gather qualitative insights into electrochemical reactions.

    Advantages of Cyclic Voltammetry

    • Provides a quick way to pinpoint redox potentials of electroactive species.
    • Can offer valuable kinetic details about electrode reactions.
    • Electrode reactions involve electron transfer influenced by electrode potential.
    • Mass transfer (diffusion) may sometimes control the overall electrode reaction.

    Disadvantages of Cyclic Voltammetry

    • The method cannot differentiate between slow heterogeneous electron transfer and chemical reactions. When both are present, calculating rate constants requires simulation methods.
    • Throughout the experiment, there is a background charging current of magnitude nCal (where Cai is the interface capacitance at the working electrode). This limits the detection threshold to around 105 M. Furthermore, the ratio of peak faradaic current to charging current decreases with increasing n, placing an upper limit on n values that can be utilized.

    Overall, cyclic voltammetry is a technique that controls the electrode's potential while measuring the resulting current.

    Application

    • Cyclic voltammetry is a technique used to analyze electrochemical processes under different conditions, providing insights into intermediates in oxidation-reduction reactions and reaction reversibility.
    • Cyclic voltammetry helps determine electron stoichiometry, analyte diffusion coefficient, and formal reduction potential, serving as an identification tool.
    • Concentration of an unknown solution can be found by creating a current vs. concentration calibration curve due to the proportional relationship between concentration and current in a reversible, Nernstian system.

    Questions - Part A

    1. Explain Beer-Lambert's law.
    2. What are the limitations of Beer-Lambert's law?
    3. Define UV-Visible spectrophotometer.
    4. Describe the sources of UV and visible light in a UV-Visible spectrophotometer.
    5. Discuss the requirements of solvents used in a UV-Visible spectrophotometer.
    6. Provide two applications of UV-Visible spectrophotometry.
    7. Explain the principle of an IR spectrophotometer.
    8. List the detectors used in an IR spectrophotometer.
    9. Give two applications of an IR spectrophotometer.
    10. Explain the principle of thermogravimetry.
    11. Define the principle of DTA (Differential Thermal Analysis).
    12. Elaborate on the principle of DSC (Differential Scanning Calorimetry).
    13. List two applications of TG (Thermogravimetry).
    14. Discuss two applications of DTA.
    15. Explain the principle of DSC and provide two applications.
    16. Define sensors and mention their types.

    Paraphrased Explanation

    Application of Cyclic Voltammetry

    • Cyclic voltammetry is a valuable method for investigating electrochemical processes, shedding light on reaction intermediates and reversibility.
    • It aids in determining electron stoichiometry, analyte diffusion coefficient, and formal reduction potential, offering identification capabilities.
    • By establishing a calibration curve of current versus concentration, one can ascertain the concentration of an unknown solution based on the Nernstian system's proportional relationship between concentration and current.

    Test Questions - Part A

    1. Explain the concept of Beer-Lambert's law with examples.
    2. Discuss the constraints associated with Beer-Lambert's law.
    3. Define a UV-Visible spectrophotometer and its operational significance.
    4. Elaborate on the light sources utilized in UV-Visible spectrophotometers.
    5. Outline the essential solvent requirements in UV-Visible spectrophotometry.
    6. Illustrate two practical applications of UV-Visible spectrophotometry.
    7. Clarify the fundamental principle underlying IR spectrophotometry.
    8. Identify the types of detectors commonly employed in IR spectrophotometry.
    9. Provide examples of two applications of IR spectrophotometry.
    10. Elucidate the operational principle of thermogravimetry and its significance.
    11. Define the principle of DTA and its role in analytical chemistry.
    12. Explain the principle of DSC and its applications in material science.
    13. Highlight two practical applications of TG for material analysis.
    14. Explore two specific applications of DTA in thermal analysis.
    15. Detail the principle of DSC, showcasing its significance in thermal analysis, and present two applications.
    16. Define sensors and categorize them based on their functionality.

    Key Concepts in Analytical Chemistry

    • Different Types of Oxygen Sensors

      In the realm of analytical chemistry, various oxygen sensors are utilized to measure oxygen levels in different environments. These sensors include electrochemical sensors, optical sensors, and solid-state sensors.

    • Applications of Oxygen Sensors

      Oxygen sensors find extensive applications in industries such as automotive, healthcare, environmental monitoring, and aerospace. They are crucial for ensuring safety, efficiency, and quality in various processes.

    • Distinguishing DSC and DTA

      DSC (Differential Scanning Calorimetry) and DTA (Differential Thermal Analysis) are analytical techniques used to study thermal properties of materials. DSC measures heat flow, while DTA measures temperature differences between a sample and a reference.

    • Understanding Glucose Sensors

      A glucose sensor is a biosensor that detects and measures glucose levels in biological systems, such as blood or interstitial fluid. These sensors are vital for monitoring and managing diabetes.

    • Defining Cyclic Voltammetry

      Cyclic voltammetry is an electrochemical technique that measures the current response of an electrochemical cell to a varying applied potential. It provides insights into the redox properties of compounds.

    • Advantages of Cyclic Voltammetry

      Cyclic voltammetry offers several advantages, including high sensitivity, rapid data acquisition, and the ability to study electrochemical reactions in situ. These characteristics make it a valuable tool in electrochemistry research.

    • Applications of Cyclic Voltammetry

      Cyclic voltammetry is employed in various fields such as material science, environmental monitoring, corrosion studies, and the development of electrochemical sensors. It helps in understanding electron transfer processes and reaction mechanisms.

    Instrumentation and Principles of Analytical Techniques

    • Laws of Absorption and Beer-Lambert's Law

      The laws of absorption govern the interaction of light with matter. Beer-Lambert's law relates the absorption of light to the properties of the material through which it passes. It is expressed mathematically as A = εlc, where A is absorbance, ε is molar absorptivity, l is path length, and c is concentration.

      Limitations of Beer-Lambert's law include assumptions of monochromatic light, dilute solutions, and a linear relationship between absorbance and concentration.

    • UV-Visible Spectrophotometer Instrumentation

      A UV-Visible spectrophotometer measures the absorption of ultraviolet and visible light by a sample. It consists of a light source, monochromator, sample holder, and detector. A block diagram illustrates the components and flow of light through the instrument.

    • IR Spectrophotometer Instrumentation

      An IR spectrophotometer analyzes the interaction of infrared radiation with molecules in a sample. It comprises a radiation source, interferometer, sample compartment, and detector. A block diagram visually represents the setup of an IR spectrophotometer.

    • Thermogravimetry Principles and Applications

      Thermogravimetry is a technique used to study changes in a sample's weight as a function of temperature. It helps in understanding decomposition, stability, and composition of materials. Applications include quality control, material characterization, and pharmaceutical analysis.

    • Differential Thermal Analysis Fundamentals

      Differential thermal analysis measures temperature differences between a sample and a reference material as they are subjected to controlled temperature changes. It is crucial for studying phase transitions, purity, and thermal behavior of substances.

    • Differential Scanning Calorimetry Overview

      Differential scanning calorimetry is a thermal analysis technique that measures heat flow in a sample as a function of temperature. It provides insights into phase transitions, heat capacity, and material properties. DSC is widely used in materials science, pharmaceuticals, and polymer research.

    Understanding Sensors and Electrochemical Techniques

    • Overview of Sensors

      Sensors are devices that detect and respond to physical or chemical stimuli, converting them into measurable signals. They play a crucial role in various fields, including healthcare, environmental monitoring, and industrial processes.

    • Types of Oxygen Sensors

      Oxygen sensors can be classified into electrochemical sensors, optical sensors, and solid-state sensors. Each type operates based on different principles but aims to measure oxygen concentrations accurately.

    • Exploring Biosensors and Glucose Sensors

      Biosensors are analytical devices that incorporate biological components to detect specific analytes. A glucose sensor, for instance, is a biosensor designed to measure glucose levels in biological samples. These sensors are essential tools in medical diagnostics and monitoring.

    • Working Principle of Cyclic Voltammetry

      Cyclic voltammetry involves applying a varying potential to an electrochemical cell and measuring the resulting current. This technique allows researchers to investigate redox reactions, electrode processes, and electrochemical kinetics with high sensitivity and precision.

    Educational Content Creation

    Content Creation Specialization

    • Proficient in summarizing and synthesizing complex information from PDF books into concise, comprehensible notes and test materials.
    • Expertise in extracting key theoretical concepts and providing clear definitions.
    • Skilled in offering illustrative examples to enhance understanding.

    Content Formatting Expertise

    • Specialize in converting text into well-formatted HTML elements.
    • Capable of structuring text using HTML tags like <li>, <ul>, and <p>.
    • Ensure that the HTML output is organized and visually appealing.

    Test Question Creation

    • Create informative and engaging test questions that accurately assess understanding of the material.
    • Optimize the learning experience by explaining complex content in digestible formats.
    The document Instrumental Techniques | Chemistry for EmSAT Achieve is a part of the EmSAT Achieve Course Chemistry for EmSAT Achieve.
    All you need of EmSAT Achieve at this link: EmSAT Achieve
    195 videos|213 docs|157 tests

    Top Courses for EmSAT Achieve

    195 videos|213 docs|157 tests
    Download as PDF
    Explore Courses for EmSAT Achieve exam

    Top Courses for EmSAT Achieve

    Signup for Free!
    Signup to see your scores go up within 7 days! Learn & Practice with 1000+ FREE Notes, Videos & Tests.
    10M+ students study on EduRev
    Related Searches

    past year papers

    ,

    Previous Year Questions with Solutions

    ,

    shortcuts and tricks

    ,

    study material

    ,

    Instrumental Techniques | Chemistry for EmSAT Achieve

    ,

    MCQs

    ,

    Objective type Questions

    ,

    Instrumental Techniques | Chemistry for EmSAT Achieve

    ,

    Free

    ,

    pdf

    ,

    Instrumental Techniques | Chemistry for EmSAT Achieve

    ,

    Exam

    ,

    video lectures

    ,

    Sample Paper

    ,

    mock tests for examination

    ,

    Summary

    ,

    practice quizzes

    ,

    Semester Notes

    ,

    Important questions

    ,

    ppt

    ,

    Viva Questions

    ,

    Extra Questions

    ;